Perform the Z-port Error Corrections
1. Connect adapter A3 to adapter A2 on port 2. (See Figure 5-12.)
2. Perform the 2-port error correction using calibration standards appropriate for the
connector type at port 1.
Note
When using adapter removal calibration, you must save calibration sets to the
internal disk, not to internal memory.
Caution
Do not mistake the Iine switch for the disk eject button. See the figure below.
If the Iine switch is mistakenly pushed, the instrument wiII be turned off,
losing all settings and data that have not been saved.
D I S K E J E C T
L I N E SWITCH/
3. Save the results to disk. Name the file "PORTl."
5 4 2
Optimizing Measurement Results
N E T W O R K ANAL'IZER
REFERENCE
PORT 1
Figure 5-12. 'lko-Port Cal Set 1
B U T T O N 1
REFERENCE