Start of Sequence
FILE NAME
PLOT NAME
SINGLE
SAVE FILE 0
PLOT
DECR LOOP COUNTER
IF LOOP COUNTER 0 THEN DO
SEQUENCE 2
Sequence 1 initializes the loop counter and calls sequence 2. Sequence 2 repeats until the loop
counter reaches 0. It takes a single sweep, saves the data file and plots the display.
The data file names generated by this sequence will be:
n
through
The plot llle names generated by this sequence will be:
n
through
'lb nm the sequence, press:

Limit Test Example Sequence

This measurement example shows you how to create a sequence that will branch the sequence
according to the outcome of a limit test. Refer to "'lksting a Device with Limit Lines," located
earlier in this Chapter, for a procedure that shows you how to create a limit test.
For this example, you must have already saved the following in register 1:
a series of active (visible) limit lines
n
an active limit test
n
1. To create a sequence that will recall the desired instrument state, perform a limit test, and
branch to another sequence position based on the outcome of that limit test, press: