8. Tbmaximiz e the dynamic range in the stopband (increasing the incident power and
narrowing the IF bandwidth), press:
. . . , . : . : . : . . . . . . < : ; . . , : , : : : : : : . y . . : . . . : . : . . : : . <
i

Calibrate and Measure

1.
Remove the DUT and perform a full two-port calibration. Refer to Chapter 5, "Optimizing
Measurement Results. n
With the thru connected, set the scale to autoscale to observe the benefits of using swept
2.
list mode.
The segments used to measure the stopbands have less noise, thus maxinUng dynamic
n
range within the stopband frequencies
The segment used to measure the passband has been set up for faster sweep speed with
n
more measurement points.
Figure Z-52. Calibrated Swept List Thru Measurement
Reconnect the filter and adjust the scale to compare results with the first filter measurement
3.
that used a linear sweep.
In Figure 2-53, notice that the noise level has decreased over 10 dB, conllrming that the
n
noise reduction techniques in the stopbands were successful
In Figure 2-53, notice that the stopband noise in the third segment is slightly lower than in
n
the first segment. This is due to the narrower IF bandwidth of the third segment (300 Hz).
i