Accessories Supplied; Other Options - HP 4262A Operating And Service Manual

Lcr meter
Hide thumbs Also See for 4262A:
Table of Contents

Advertisement

Section I
Paragraphs 1-17 to 1-33
1-17.
For information concerning a serial number
prefix that is not listed on the title page or in the
Manual
Changes supplement, contact
your. nearest
Hewlett-Packard office.
1-18.
OPTIONS.
1-19.
Options for the Model
4262A
LCR Meter
are available for adding the following capabilities:
Option 001: BCD Parallel Data Output.
Option 004: Comparator.
A comparator func-
tion providing GO/NO-GO judge-
ment with HIGH and LOW limits
for LCR and D/Q.
Option 101: HP-IB Interface.
Option 010: 100Hz Test Frequency.
(instead of 120Hz)
_ Options 907, 908 or 909 are handle or rack
__
mount
kits. See paragraph 1-29
for details.
Option 910: Extra Manual.
1-20.
OPTION 001.
1-21.
The 4262A
option 001 provides separate
BCD
parallel data output
for L, C, R/ESR
and
dissipation factor or quality factor simultaneously
from the two rear panel connectors. With this op-
tion,
external
data processing
devices
such as a
digital printer can be used
with
the 4262A.
1-22. OPTION 004.
1-23.
The
4262A
Option
004
provides
for
GO/NO-GO judgement by comparing L, C, R/ESR
and D/Q values to HIGH
and LOW
limits. Three
judgement
outputs are provided:
LED
lamp dis-
play, relay contacts, or TTL
level voltages (open
collectors):
HIGH
. .measured value is not less than HIGH
limit.
IN... measured value is less than HIGH limit
and not less than LOW limit.
LOW ...measured value is less than LOW limit.
Model 4262A
1-24.
OPTION 101.
1-25.
The 4262A
Option 101 provides interfac-
ing functions
to both
transfer L, C, R/ESR
and
D/Q data to HP Interface Bus line and to receive
remote control signals from HP Interface Bus line.
1-26.
OPTION 010.
1-27.
The
4262A
Option 010 provides test fre-
quencies
of 100Hz,
1kHz, and 10kHz
(100Hz
is
used instead of standard 120Hz). All other electri-
cal performance is the same as that of standard in-
strument.
1-28.
OTHER OPTIONS.
ing:
Option 907:
Front Handle Kit.
Option 908:
Rack Flange Kit.
1-31.
ACCESSORIES SUPPLIED.
1-32.
Figure
1-1 shows
the HP Model 4262A
LCR Meter, power cord (HP Part No. 8120-1378),
and fuses (HP Part No. 2110-0007 and 2110-0202),
1-33.
ACCESSORIES AVAILABLE.
1-34.
For effective and easy measurement, three
styles of fixtures and leads for the measurement
of various components
are available.
These
are
listed in Table 1-1.A brief description of each of
these fixtures and leads is given in Table 1-3. Refer
to Section III Figure 3-3 on page 3-8 for detailed
information on these devices.

Advertisement