LAN connector (RJ-45) is used to integrate the instrument to a Local Area Network,
primarily for remote control purposes. This connector is not available on the
R&S OSP150.
CAN bus connector (D-Sub, 9 pin) is a control port to connect one or several extension
units R&S OSP150.
The R&S OSP owns three slots which can be configured with the options available for
the R&S OSP. The above configuration shows the following options:
Option R&S OSP-B101: Relay module consisting of six coaxial relays of SPDT type;
the RF coaxial connectors are SMA type
Option R&S OSP-B102: Relay module consisting of two coaxial relays of SP6T type;
the RF coaxial connectors are SMA type
Option R&S OSP-B103: I/O module with 16 Bit input / output ports; the connectors are
D-Sub types, 25 pin, female for output and male for input
In case of the options which can be installed in the R&S OSP, the connectors related
to each option are described in detail in

2.4.1 LAN Connector

8-pin LAN connector RJ-45 used to connect the R&S OSP120 or OSP130 to a Local
Area Network (LAN). Refer to Remote Operation in a LAN.
LAN Connection
Depending on the connection (a non-dedicated network connection or dedicated
connection to a single controller) a standard RJ-45 cable or cross-over RJ-45 cable is
required. See
It is recommended to use double-shielded LAN cables of category 6 (SSTP).
Operating Manual 1505.3896.12 - 14
chapter 2.9.1
for further information.
chapter
9.2,
R&S OSP Module
Preparing for Use
Interfaces.