R&S Osp120 Version12 And Modules R&S Osp-B011/-B012; Figure 5-6: Block Diagram R&S Osp120 Var. 12 - R&S OSP Operating Manual

Open switch and control unit
Table of Contents

Advertisement

R&S OSP
Standard Configuration:
The slots in the front side are designated with A12F and A13F. For installation of
modules, the following order is defined as standard:
The first module installed from OSP front side always is mounted in slot A13F. From
interface point of view it corresponds to slot A13 (connected via cable W7 – see figure
below).
The second module installed from OSP front side is mounted in slot A12F. From
interface point of view it corresponds to slot A12 (connected via cable W6 – see figure
below).
Figure 5-6: Block Diagram R&S OSP120 Var. 12
Configuration of 2-slot modules:
OSP modules with 2-slot size and requiring two interface cables W6 and W7 (like the
modules OSP-B105/-B115/-B125/-B126) only can be mounted in A12/A13 or
A12F/A13F.
That means with a 2-slot module configured in A12/A13 and one module configured in
the OSP front side, only interface cable W5 (corresponding to A11) is left for A13F).
5.1.4 R&S OSP120 Version12 and Modules R&S OSP-B011/-B012
Configuring a system with instruments having the RF connectors at the front panel, the
cabling will be made easy having the RF connectors at the OSP120 front side as well.
The R&S OSP120 Version 12 is designed with two slots in the front side of the in
addition to the three slots at the rear side. Two slots therefore can be used for
mounting the options OSP-B011 and/or OSP-B012. See following diagram.
Operating Manual 1505.3896.12 - 14
Instrument Functions
79