2.5.2 Instrument Setup

The R&S OSP is designed for use under laboratory conditions, either on a bench top
or in a rack. The general ambient conditions required at the operating site are as
follows:
The ambient temperature must be in the ranges specified for operation and for
compliance with specifications (see "Specifications").
All fan openings including the rear panel perforations must be unobstructed. The
distance to the wall should be at least 10 cm.

2.5.3 Bench Top Operation

If the R&S OSP is operated on a bench top, the surface should be flat.
The instrument is used in horizontal position, standing on its feet.

2.5.4 Mounting in a 19" Rack

The instrument can be mounted in 19" racks using a ZZA-211 adapter (order number
1096.3260.00). Please note the mounting instructions supplied with the rack adapter.
Allow for sufficient air supply in the rack.
Make sure that there is sufficient space between the ventilation holes and the rack
casing.

2.5.5 EMI Protective Measures

In order to avoid electromagnetic interference (EMI), the instrument may only be
operated when it is closed and with all shielding covers fitted. Only appropriate
shielded signal and control cables may be used.
Operating Manual 1505.3896.12 - 14
Preparing for Use
21