7.4.6 SYSTem Commands

The SYSTem subsystem contains functions that are not related to instrument
performance. All commands are SCPI confirmed.
SYSTem:DFP?
Queries the device footprint of the instrument, which means the equipment's
configuration. The retrieved information is in machine-readable form (for example in
xml format), suitable for further processing.
Example
Characteristics
SYSTem:ERRor:ALL?
Queries and at the same time deletes all entries in the error queue.
The entry consists of an error number and a short description of the error. Positive
error numbers are instrument-dependent. Negative error numbers are reserved by the
SCPI standard.
Example
Characteristics
Operating Manual 1505.3896.12 - 14
"F01A13|OSP-B103|100001|1505.3270.02|0|01.00",
"F02|OSPMAINBOARD|100123|1505.3050.04|0|03.00",
"F02A11|OSP-B101|100001|1505.3250.02|0|01.00",
The hardware information query command detected two
R&S OSP instruments with ID F01 and F02. The instrument F01
comprises a mainboard and three modules:
B101 located at A11,
B102 located at A12,
B103 located at A13.
The instrument F02 comprises a mainboard and only one
module:
B101 located at A11.
Firmware version V1.0
Device-specific. Query only.
SYST:DFP?
Queries the device footprint of the instrument and returns an
ASCII string.
Firmware version V.57.
SCPI confirmed. Query only.
SYST:ERR:ALL?
Query all entries in the error queue. "0,"No error"" is returned if
the error queue is empty.
Firmware version V1.00.