R&S OSP Operating Manual page 170

Open switch and control unit
Table of Contents

Advertisement

R&S OSP
ROUTe:MODule:INIT
The INIT command causes an update of all internal held data of the system of one or
several R&S OSP instruments connected via CAN bus. The internal held data
comprises data like serial number, part number, hardware code etc. of all modules of
all R&S OSP instruments that were previously announced to the system with the
command ROUTe:MODule:DEFine <instr_name>,<instr_address>.
The data is read from the flash EEPROM of all modules.
There is no need to send the INIT command right after a new R&S OSP instrument
was defined or deleted. To define or delete an instrument means to update the
internal held data automatically.
Example
Characteristics
ROUTe:MODule:CATalog?
The ROUTe:MODule:CATalog? query command returns a list of all relay or I/O
channel modules of all R&S OSP instruments that were previously announced to the
system with the command ROUTe:MODule:DEFine. Besides the internal module
name (e.g. F01A11), also the type of the module is returned (e.g. R&S OSP-B101).
Example
Characteristics
ROUTe:CLOSe <channel_list>
The CLOSe command allows specific individual relays and I/O channels to be set. The
CLOSe? query allows the condition of individual relays and I/O channels to be queried.
The instrument returns a 1 or 0 for each relay or channel in the list, in the same order
as the list is specified. A response of 1 means the relay or the channel is in the
condition that is indicated in the <channel_list>. A response of 0 means the channel is
not in the condition that is indicated in the <channel_list>.
Operating Manual 1505.3896.12 - 14
ROUT:MOD:INIT
Reads relevant data from the flash EEPROM of all modules of
all R&S OSP instruments that build up a system.
Firmware version V1.0
Device-specific. No query.
ROUT:MOD:CAT?
Response: "OSP-B101-F01A11","OSP-B102-F01A12"
For further information on naming conventions of modules,
module locations within an instrument and instrument names,
please refer to
chapter
Firmware version V1.0
SCPI-confirmed, but device-specific response format.
Query only.
Remote Command Reference
7.2, Naming Conventions.
155