Figure 5-42: Module R&S OSP-PM-I
The module R&S OSP-PM-I just is used as housing for the Power Sensor. There is no
connection to R&S OSP internal interface. See following block diagram.
Figure 5-43: Block diagram Module R&S OSP-PM-I
The R&S OSP-PM-I is a one-slot module. Within one R&S OSP120 up to three
modules of this type can be installed.
5.2.15 RF Switch Module R&S OSP-B116 and R&S OSP-B136
The module R&S OSP-B116 and B136 are designed as a universal switching modules
using two transfer switches DPDT.
The modules are designed for independent control of both relays.
The module R&S OSP-B116 is configured with two transfer relays and SMA connector
type, the R&S OSP-B136 with two transfer relays and N connector type.
When not activated (no voltage applied to the relay) and after Reset, the terminals 1 -
2 and 3 - 4 are connected.
When activated, the terminals 1 - 3 and 2 -4 are connected.
Both the modules R&S OSP-B116/-136 module is equipped with a on board memory
to store the necessary configuration data of the module.configuration data of the
module.
Operating Manual 1505.3896.12 - 14
Instrument Functions
104