3.5.3 Remote Control Setting of the Path Configuration
Once the required path configurations are stored in the R&S OSP, they are available
by remote control applications like the R&S EMC32 software.
The advantage to make use of the OSP path configuration utility is:
●
●
For the above examples, the following strings are to be sent to the R&S OSP:
ROUTe:CLOSe "Path_PA1"
ROUTe:CLOSe "Path_PA2"
ROUTe:CLOSe "Path_PA3"
Further information on remote control functions are to be found in
Information of applications together with the R&S EMC32 software is available in
chapter
Operating Manual 1505.3896.12 - 14
The switching is called by logical name from the external application (for example
EMC32 software). Using the logical name, the external application does not need
to know the R&S OSP hardware configuration.
If the hardware configuration has to be changed, only the path configuration
inside the R&S OSP must be adapted. The external application does not need to
be changed.
8.1.
Getting Started
chapters 6
and 7.
68