R&S OSP Operating Manual page 169

Open switch and control unit
Table of Contents

Advertisement

R&S OSP
'<instr_address>'
Example - Set
Example - Query
Characteristics
ROUTe:MODule:DELete <instr_name>
The previously defined <instr_name> will be deleted from the instrument's internal
memory and from the compact flash. Afterwards the relays and I/O channels of this
instrument cannot be controlled any longer.
'<instr_name>'
Example - Set
Characteristics
ROUTe:MODule:DELete:ALL
All previously defined <module_names> will be deleted from the instrument's internal
memory and from the compact flash. Afterwards the relays and I/O channels of all
instruments that built up a system cannot be controlled any longer.
Example - Set
Characteristics
Operating Manual 1505.3896.12 - 14
Address of an R&S OSP instrument, built up from the instrument
type ID and the serial number of the instrument.
Instrument type ID:
2: R&S OSP120
3: R&S OSP130
5: R&S OSP150
Serial number: for example 100001
ROUT:MOD:DEF F01,3100001
Assigns the R&S OSP130 instrument with serial number 100001
to the instrument name F01.
ROUT:MOD:DEF? F01
Response: 3100001
Firmware version V1.0
SCPI-confirmed
Represents the name of an R&S OSP instrument, which
previously was assigned to the instrument.
Format: F01, F02, .. up to F09
ROUT:MOD:DEL F03
All relay and I/O channel modules of the R&S OSP instrument
assigned to the name 'F03' are not longer available.
Firmware version V1.0
SCPI-confirmed. No query.
ROUT:MOD:DEL:ALL
All relay and I/O channel modules of all R&S OSP instruments
cannot be controlled any longer.
Firmware version V1.0
SCPI-confirmed. No query.
Remote Command Reference
154