Toc Calibration Procedure - Hach ANATEL PAT700 Operator's Manual

On-line total organic carbon analyzer
Hide thumbs Also See for ANATEL PAT700:
Table of Contents

Advertisement

PAT700 Total Organic Carbon Analyzer – Bottle Mode
Bottle selection check boxes
The default setting is all three bottles checked.
Reps text box
may enter a value from 1 to 3. The default setting is 3.
Save as default button
Back icon
setup are retained.
Forward icon
are retained. See
Cancel button
dialog boxes are lost.

TOC calibration procedure

CAUTION
Potential puncture hazard! Contact with the exposed needle could cause serious
injury. Use extreme care when installing and removing the sample bottles.
1)
2)
3)
4)
Anatel
Figure 6-4 Custom setup dialog box
enable you to enable or disable the bottles to use in the test.
allows you to specify the number of repetitions to perform on each bottle. You
saves the bottle selection and reps as the default setting.
returns you to the TOC calibration setup dialog box. All changes made to custom
takes you to the load bottles dialog box. All changes made to the custom setup
Figure
6-5.
takes you to the run standards dialog box. All changes made in any of the
Allow the TOC calibration standards to warm to room temperature.
At the home screen, touch the << icon to open the sliding toolbar, then touch the
bottles icon in the toolbar, or touch the bottles icon in the lower left corner of the
screen.
Touch the run standards button in the bottle mode dialog box. See
page
76.
Touch the TOC calibration button in the run standards dialog box. See
page
77.
Page 79
Figure 6-1,
Figure 6-2,
Operator Manual