Inflation Pressure Default Setting; Procedures; Alarm Default Settings; Spo2 Configuration Settings - GE DINAMAP ProCare Service Manual

Table of Contents

Advertisement

3-8
Installation: Configuring Your ProCare Monitor
Inflation Pressure Default Setting –

Procedures

1.
Enter the configuration mode: with the Monitor off, press and hold the Menu
button at the same time as pressing and holding the On/Off button for 3
seconds, or press Menu until the Inflate Pressure is lit on the display and the
pressure is displayed in the Diastolic window.
2.
Use the +/- buttons to increment or decrement the inflate pressure default
setting. Increments are 5mmhg from 100mmhg to 250mmhg.
3.
To exit the configuration mode, turn the unit off. To continue with additional
configuration settings, press Menu.
Alarm Default Settings –
Procedures
1.
Enter the configuration mode: with the Monitor off, press and hold the Menu
button at the same time as pressing and holding the On/Off button for 3
seconds. After the unit enters the configuration mode, press Alarms. At any
point in the configuration mode menu, Alarms default can be selected.
2.
To set or change the default setting, press the Alarms button to select alarm
setting. Use the +/- buttons to increment or decrement the individual
settings.
NOTE: For the Alarms default setting to be saved, you must advance the
menu through the SPO2 settings.
3.
To exit the configuration mode, turn the unit off. To continue with additional
configuration settings, press Menu.
SpO
Configuration Settings –
2
Procedure for units with Nellcor® Technology
(Refer to the Nellcor® section for options)
1.
Enter the configuration mode: with the Monitor off, press and hold the Menu
button at the same time as pressing and holding the On/Off button for 3
seconds.
2.
Press the menu button until
3.
Use the +/- buttons to select the option.
4.
Press the Menu button once. SAt appears in the Pulse Rate window.
5.
Use the +/- buttons to select the option.
6.
To exit the configuration mode, turn the unit off. To continue with additional
configuration settings, press Menu.
®
DINAMAP
ProCare Monitor Service Manual
2009381-001
n
0d appears in the Pulse Rate window.
Revision C