MP8 Interface Emulation

Purpose
With the MP8 interface emulation you
can connect peripheral devices of the
MP8 generation that have separate AC
power supplies, such as the 73822...
Data Control terminal, a YFC... Flow
Rate Controller, a YDI50Z Data Input
dedicated keyboard, etc., to your GPC65
series balance.
Available Features
– The balance can only be used to deter-
mine weights
– The interface communicates exclusively
in the MP8 binary protocol.
– Select application programs for use with
the MP8 under item 3 in the balance
operating menu.
– The Index 2 program for MP8 can be
selected under item 4 of the balance
operating menu
– The following parameters remain acces-
sible as before:
– Weighing parameters [1-x-x]
– Extra functions [8-x-x]
– Reset function [9-x-x]
(see "Setting Parameters (Menu)" in
the chapter entitled "Configuration")
"MP8 interface" not allowed during use
!
in legal metrology. When the menu
access switch is sealed, the MP8 inter-
face will not function.
Factory Settings of the Parameters
(special settings for MP8 functions)
Program selection: MP8: 3-1-1
Program index 2: 1 Ind. 2.1
Preparation
Turn on the balance: Press e
> All display segments light up briefly
Switch to the MP8 interface:
Press M
Select balance operating menu: Menu soft key (press the w key)
Select and confirm the balance menu: press the y and then the X key
Confirm factory setting: X key
Select and confirm MP8 mode [9-1-9]
Press y or Y key, repeatedly if necessary; then press X
Press M
Parameter Settings for the MP8 Interface
Press M
Select the balance operating menu: Menu soft key (press the w key)
Select and confirm:
3 Application prog. 1 app. selection:
1 MP8 3-1-1 or
...
9 MP8 3-1-9 or
10 MP8 3-2-1 or
...
18 MP8 3-2-9 or
19 MP8 3-3-1 or
...
27 MP8 3-3-9
4 Program-Index; 2 Ind. 2:
1 Ind. 2.1 or
2 Ind. 2.2 or
3 Ind. 2.3 or
4 Ind. 2.4
5 Interface:
1 Baud rate
1 150 Baud or
2 300 Baud or
3 600 Baud or
4 1,200 Baud or
5 2,400 Baud or
6 4,800 Baud or
7 9,600 Baud or
2 Parity
2 Space or
3 Odd or
4 Even
6 Print weighing
1 print Man./auto
1 Manual without stability or
2 Manual with stability or
4 Automatic without stability or
5 Automatic at stability
§ Store settings and exit the Setup menu: Press M