The Drystar 5302 will continue with the 'Select for Modality' screen, allowing
5
you to change the Look-up table, the Interpolation and the maximum density
(Dmax) values manually.
<Modality name>
3 Dmax
In the 'Select for Modality' screen, press the Down key, followed by the Confirm
6
key to select 'Interpolation'.
The 'Interpolation' screen is displayed:
<Modality name>
1*Replication
2 Linear
3 Cubic Hi-res
4 CubicBel
5 None
Press the up/down key to select the desired interpolation setting for this
7
modality.
If you select "Cubic Hi-Res" the following screen will appear, allowing you to
adjust the smoothing factor:
Enter the smoothing factor value by means of the arrow keys. Refer to
on page
31
(valid range [-5.00, 0.00]).
If you select "CubicBell", the following screen will appear, allowing you to
adjust the smoothing factor:
Enter the smoothing factor value by means of the arrow keys. Refer to
on page
31
(valid range [0.10, 1.00]).
Press the Confirm key to store the data, or the Escape key to quit the procedure
8
without any changes.
The user should be aware that the interpolation values differ when the
A#Sharp feature is deactivated. The A#Sharp feature, which is a
technology that enhances the image sharpness, is active by default. For
more information about (de)activating the A#Sharp feature, refer to the
Drystar 5302 Service manual.
2830A EN 20050526
D
5302 R
RYSTAR
EFERENCE MANUAL
'Data entry'
'Data entry'
Advanced operation (Key-operator mode)
95