AGFA Drystar 5302 Reference Manual page 90

Hide thumbs Also See for Drystar 5302:
Table of Contents

Advertisement

D
5302 R
RYSTAR
EFERENCE MANUAL
Press the up/down key to select the preferred film calibration default.
5
When the setting is ON, the Drystar 5302 will ask you to enter the calibration
frequency:
Enter the calibration frequency by means of the up/down key. Refer to
page 31.
1 pack = 100 sheets
E.g., if the frequency is set to 5, this means that a film calibration is executed every
5 packs or every 500 sheets.
You can specify to perform an automatic film calibration after loading every "x"
film packs.
For more information on film calibration settings: refer to
calibration policy'
Press the Confirm key to store the data, or the Escape key to quit the procedure
6
without any changes.
90
Advanced operation (Key-operator mode)
Automatic film calibration after every 'x' packs
on page 195.
'Data entry'
on
'Understanding the
2830A EN 20050526