Press the Up/Down arrow keys to select the desired tray, followed by the
5
Confirm key.
The following screens appear successively:
The Drystar 5302 will automatically print the QC test image.
After the image is printed, the system will display the optical density values:
6
DD: 1.56
(Densitometer=MB924)
The displayed values represent the following steps on the test film:
Low density
Mid density
High density
Base + Fog,
Density difference
(DD),
Maximum density
(Max D)
If the mid density value does not meet or exceeds the recommended
values, the reason must be found and the problem solved before any
further clinical films can be printed.
Refer to
quality and resolving image quality problems'
Agfa service organization.
2830A EN 20050526
Operating Level
the density value of the low
density step
the density value of the mid
density step
the density value of the high
density step
These values are displayed but not important for this QC
procedure
'Preventive maintenance schedule'
D
RYSTAR
The density levels are displayed, e.g:
Max D: 3.1
Hi D: 1.91
Mid D: 1.13
Lo D: 0.35
Base+Fog: 0.19
DD: 1.56
Value (Macbeth units)
(according to
IEC 1223-2-4 or better)
0,4 ± 0,05
1,2 ± 0,15
2,0 ± 0,2
on page
172
and
on page 194, or call your local
Advanced operation (Key-operator mode)
5302 R
EFERENCE MANUAL
'Maintaining image