Network connectivity
Ethernet / connectors
Network protocols (TCP/IP services)
Image formats
Postscript
Power consumption - heat dissipation
During operation
In stand-by
Peak power (absolute max. rating)
Protection against
Electrical shocks
Ingress of water
Environmental conditions (operation)
Room temperature
Relative humidity
Atmospheric pressure
Environmental storage conditions
Climate conditions for storage are in accordance with EN60721-3-1-class 1K4.
Room temperature
Relative humidity
Absolute humidity
Rate of change of temperature
Atmospheric pressure
2830A EN 20050526
D
RYSTAR
RJ45 twisted pair for 10/100Base-TX;
Serial RS232 connection
FTP, Telnet, HTTP
DICOM (Default)
TIFF
Not available
250 W - 900 kJ/h
70 W - 252 kJ/h
530 W - 1908 kJ/h
Class 1 (grounded)
IPXØ
Between +15°C and +30°C
Between 20% and 75%
Note: Films may not become wet!
70 kPa - 106 kPa
Between -25°C and 55°C (storage)
Between 10% and 100%
3
Between 0.1 g/m
and 35 g/m
1°C/min
70 kPa - 106 kPa
5302 R