Performing quality control (QC) tests
The following procedures must be performed daily, weekly or annually as
indicated.
The reason for performing quality control tests is to determine if any
significant image quality variation or deterioration has occurred which may
require corrective action. Comparing the results of the tests with the reference
values previously established does this.
This procedure allows the operator to take the necessary preventive actions
before any image quality loss can take place.
Performing the daily QC test
This test must be performed every day before any clinical film can be
processed.
Turn on the Drystar 5302 and wait at least for 15 minutes. Refer to
1
the Drystar 5302'
Press the Key-operator key to enter the Key-operator mode.
2
Press the down key seven times, followed by the ok key to select 'QC'.
3
5 Restore config.
6 Calibration
7 Service Actions
8 Quality Control
9 Installation
A confirmation screen appears:
Do one of the following:
4
Press the Confirm key (YES) to start the QC procedure and proceed with
step 5.
Press the Escape key (NO) to quit.
2830A EN 20050526
on page 32.
D
5302 R
RYSTAR
EFERENCE MANUAL
'Switching on
Advanced operation (Key-operator mode)
133