AGFA Drystar 5302 Reference Manual page 212

Hide thumbs Also See for Drystar 5302:
Table of Contents

Advertisement

D
5302 R
RYSTAR
EFERENCE MANUAL
Drystar 5302: Determination of Operating Levels
Imager Type: __________
Film Type: ____________
Step 1: Print QC Test images on five consecutive days. Record the optical densities
measurements in the tables below. After five days, average the values to determine
the operating (aim) levels for each of the parameters.
Month
Day
Initials
Low Density
Mid Density
High Density
Average of 5 Values =
Step 2: Copy the calculated reference levels to Charts 2A/B ('Daily Density Control Chart')
212
Quality Control Charts
Serial #: _________________
Emulsion #: ______________
Densitometer: _________________ (default selection)
Day 1
Average of 5 Values = calculated reference "Low Density" level
Average of 5 Values =
calculated reference "Mid Density" level
calculated reference "High Density" level
Day 2
Day 3
Chart 1
Date ____________________
Day 4
Day 5
2830A EN 20050526