Module Connection Cables; Sensors; Status Displays - Invacare Twister Service Manual

Wheelchairs
Hide thumbs Also See for Twister:
Table of Contents

Advertisement

Twister/Twister Junior
4.2

Module connection cables

The module connection cables are called
BUS cables. They are designed as four-
conductor cables. Two conductors serve the
power supply and two the transfer of data.
Cable allocation
Power supply
= black ground-
cable
red positive cable
Data transfer
= yellow cable
(CAN high)
purple cable
(CAN low)
4.3

Sensors

The sensor unit consists of the sensor and a
magnet. If the magnet is closer than 1.5 cm
to the sensor tip, no electrical energy can
flow through the sensor.
In a vehicle with lifter, information is passed
using the sensor unit to the lifter positioning
control and the seat tilting.
4.4

Status displays

Status displays are LEDs that point to errors in the
wheelchair's electrical installation by means of blinking.
The following joystick boxes have a status display:
• joystick box (standard)
• chin control
• table and central control
The ON/OFF LED's are designed as status displays. The
causes of trouble are displayed by a blinking interval (blinking
code) of these LED's.
The remaining modules are also equipped with status
displays. A blinking of these displays points to a module
error.
Service Manual
BUS cable
BUS
Sensor unit
Magnet
Sensor
Table central control
Page 28