Twister / Twister Junior
Fault
No voltage?
Voltage below 19 V.
Note:
In the joystick box for attendant (additional joystick box, reversible) the following
inspections must be performed.
Check, whether the control can be changed over using the flip switch.
Check whether the final speed can be preselected using the control device.
Check, whether the wheelchair can be controlled via the joystick.
Check, whether all further functions (lighting and actuator functions) continue to be
controllable via the main joystick box.
Note:
Please supply the part number and the software version number when ordering a
replacement joystick box.
The software version number can be found on a sticker on the joystick box.
Issued: 21.01.02
Cause
- BUS cable defective.
- BUS plug defective.
- power module
defective.
- Completely discharge
battery
Page 165
Service Manual
Remedy
⇒ - Check/replace BUS cable. (Chapt. 19
- Check/replace BUS cable. (Chapt. 19
- Check power module. (Chapt. 18.3) -
Check power supply to power module
(Chapt. 18.3)
⇒ Check battery. (Chapt. 14.1)