Calibration Procedures
7.6. Manual Calibration Checks with IZS or Zero/Span
Valves
Zero and span checks using the zero/span valve or IZS option are similar to that
described in Section 7.3, except:
On units with an IZS option installed, zero air and span gas are supplied to the
analyzer through the zero gas inlet and from ambient air.
On units with a zero/span valve option installed, zero air and span gas are
supplied to the analyzer through the zero gas and span gas inlets from two
different sources.
The zero and calibration operations are initiated directly and independently with
dedicated keys CALZ and CALS.
To perform a manual calibration check of an analyzer with a zero/span valve or IZS
Option installed, use the following method
STEP ONE: Connect the sources of Zero Air and Span Gas as shown below.
MODEL 700
Gas Dilution Calibrator
(with O
generator option)
3
MODEL 701
Calibrated
SO
gas
2
(At high
concentration)
Figure 7-3:
136
Source of
SAMPLE Gas
VENT
External Zero
Air Scrubber
Zero Air
Generator
Source of
SAMPLE Gas
Ambient
Air
Setup for Manual Calibration Check with Z/S Valve or IZS Option
Model 100E Instruction Manual
Zero/Span Valves – Option 50
VENT if input is pressurized
Sample
Exhaust
Span
Zero Air
Filter
Internal Zero/Span Option (IZS) – Option 51
VENT if input is pressurized
Sample
MODEL
Exhaust
100E
Span
Zero Air
MODEL
100E
045150102 Rev XB1