Calibration Checks With Izs Or Zero/Span Valves; Figure 7-4: Setup For Manual Calibration Check With Z/S Valve Or Izs Option - Teledyne 200E Instruction Manual

Chemiluminescence nitrogen oxides analyzer
Table of Contents

Advertisement

Model 200E Instruction Manual
Analyzer enters SPAN CAL
mode.
If the SPAN button does
not appear, see Section
11 for troubleshooting.
Turn off the CAL_ON_NO2 feature (step four above).
7.6. Calibration Checks with IZS or Zero/Span
Valves
To perform a manual calibration check with zero/span valve or IZS option installed:
Connect the sources of Zero Air and Span Gas as shown below.

Figure 7-4: Setup for Manual Calibration Check with Z/S Valve or IZS Option

044100102 Rev A
SAMPLE
RANGE = 500.0 PPB
< TST
CAL
CALZ CALS
TST >
SAMPLE
NOX STB=XXX.X PPB
< TST TST > CAL CALZ
CALS
SPAN CAL M
NOX STB=XXX.X PPB
< TST TST >
SPAN
CONC
NOX STB=XXX.X PPB
SPAN CAL M
< TST TST >
ENTR
CONC
Pressing ENTR changes the
calibration of the instrument.
SPAN CAL M
NOX STB=XXX.X PPB
< TST TST
> SPAN
CONC
Calibration Procedures
NOX= X.X
Set the Display to show the NOX
STB test function.
SETUP
NOX=X.X
X
SETUP
Wait until the STABIL value
falls below 0.5 ppb. This
may take several minutes.
NOX=X.X
EXIT
NOX=X.X
EXIT terminates the
calibration process and
EXIT
returns the unit to
SAMPLE mode.
NOX=X.X
EXIT returns to the
SAMPLE mode.
EXIT
119