4
Troubleshooting

To verify basic oscilloscope operation

72
1 Press the Save/Recall key on the front panel, then press
the Default Setup softkey under the display. The oscilloscope
is now configured to its default settings.
2 Connect an oscilloscope probe from channel 1 to the Probe
Comp signal terminal on the front panel.
3 Connect the probe's ground lead to the ground terminal
that is next to the Probe Comp terminal.
4 Press AutoScale.
5 You should see a waveform on the oscilloscope's display
similar to this:
If you see the waveform, but the square wave is not
shaped correctly as shown above, perform the procedure
"To compensate the analog probes"
If you do not see the waveform, ensure your power source
is adequate, the oscilloscope is properly powered- on, and
the probe is connected securely to the front- panel analog
channel input BNC and to the Probe Comp terminal.
on page 73.
7000 Series Oscilloscopes Service Guide