Figure 10 User Calibration Cable For 4-Channel Oscilloscope - Agilent Technologies InfiniiVision 7000 Series Service Manual

Hide thumbs Also See for InfiniiVision 7000 Series:
Table of Contents

Advertisement

To Channel 1
To Channel 2
7000 Series Oscilloscopes Service Guide
BNC(f)- to- BNC(f) (barrel connector) to the tee as
shown below.
User Calibration cable for 4-channel oscilloscope
Figure 10
3 Connect a BNC cable (40 inches maximum) from the TRIG
OUT connector on the rear panel to the BNC barrel
connector.
4 Press the Utility key, then press the Service softkey.
5 Begin the Self Cal by pressing the Start User Cal softkey.
6 When the User Cal is completed, set the rear- panel
CALIBRATION switch to PROTECTED.
Calibrating and Adjusting
To Channel 4
3
To Channel 3
Longer cable
to TRIG OUT
63