To Remove Acquisition And Inverter Assemblies; Figure 21 Removing The T6 Screws - Agilent Technologies InfiniiVision 7000 Series Service Manual

Hide thumbs Also See for InfiniiVision 7000 Series:
Table of Contents

Advertisement

5
Replacing Assemblies

To remove Acquisition and Inverter assemblies

N O T E
90
The following illustrates how to remove the Acquisition and
Dual inverter printed circuit boards.
If the battery on the acquisition board needs to be replaced, use a
CR2477/1HF or CR2477/HFN 3V manganese dioxide lithium battery.
1 Using a TORK T6 driver locate and remove the 4 screws
on the front of the instrument (4 Channel version).
Removing the T6 screws
Figure 21
2 Disconnect Inverter, keyboard and display cables. Note
locations for re- connection. It should be noted that cables
can be removed from cable clamps at this time as well.
The cable restraining pads that affix the display cable to
the front deck are adhesive and great care should be
taken when removing them so as not to damage the cable.
7000 Series Oscilloscopes Service Guide