Carescape Ecg Specifications; Carescape Spo - GE CARESCAPE ONE User Manual

Hide thumbs Also See for CARESCAPE ONE:
Table of Contents

Advertisement

CARESCAPE ONE User Manual DRAFT 26 October 2017

CARESCAPE ECG specifications

Length
Weight
Operating temperature range
Non-operating temperature
range
Operating humidity range
Non-operating humidity range
Operating altitude range
Non-operating altitude range
Degree of enclosure protection
against solid objects and water
Power consumption
Power requirement - input
voltage
Power requirement - input
current
1 m drop
Shock
Broad-band random vibration

CARESCAPE SpO

CARESCAPE SpO
Length
Weight
Operating temperature range
Non-operating temperature
range
Operating humidity range
Non-operating humidity range
Operating altitude range
Non-operating altitude range
Degree of enclosure protection
against solid objects and water
Power consumption
2094480-001
3.7 m or 1.9 m (12.1 ft or 6.2 ft)
Less than 0.57 kg (1.26 lb), includes long 10 leadwire set
0°C to 35°C (32°F to 95°F)
-30°C to 70°C (-22°F to 158°F)
5% to 95% RH (non-condensing)
5% to 95% RH (non-condensing)
-500 m to 4000 m (1075 hPa to 616 hPa)
-500 m to 5573 m (1075 hPa to 500 hPa)
IP47
625 mW maximum
5 VDC ± 0.25 VDC
125 mA maximum
Drop from 1m in multiple orientations onto a hard surface
per the test in IEC 60601-1 clause 15.3.4.1
Withstands shock in accordance with IEC 60068-27:2008
using conditions set in IEC 60721-4-7:2001 class 7M3
Withstands broad-band random vibration in accordance
with IEC 60068-2-64:2008 using conditions set in IEC
60721-4-7:2001, class 7M3
specifications
2
(TruSignal)
2
3.0 m or 1.8 m (9.8 ft or 5.9 ft)
less than 0.17 kg (0.38 lb)
0°C to 40°C (32°F to 104°F)
-30°C to 70°C (-22°F to 158°F)
5% to 95% RH (non-condensing)
5% to 95% RH (non-condensing)
-500 m to 4000 m (1075 hPa to 616 hPa)
-500 m to 5573 m (1075 hPa to 500 hPa)
IP47
375 mW maximum
CARESCAPE ONE
Design, environmental, and physical specifications
255