GE CARESCAPE ONE User Manual page 271

Hide thumbs Also See for CARESCAPE ONE:
Table of Contents

Advertisement

CARESCAPE ONE User Manual DRAFT 26 October 2017
Alarm
Device overheating
Move Temp device to a cooler
location
ECG device failure: Call service
ECG device overheated.
Shutting down
ECG measurements removed
Identical ECG device
Identical pressure devices
Identical SpO2 device
Identical Temp device
Incompatible probe
Incompatible SpO2 probe
Incompatible adhesive SpO2
probe
Incompatible SpO2 device
[Invasive pressure channel label]
X sensor disconnected, where
[Invasive pressure channel label]
= Art, CPP, CVP, Fem, FemV, ICP,
LAP, P, PA, RAP, RVP, UAC, or
UVC, and X = invasive pressure
channel number 1 to 2.
Sensor
[Invasive pressure channel
label] X sensor failed, where
[Invasive pressure channel label]
= Art, CPP, CVP, Fem, FemV, ICP,
LAP, P, PA, RAP, RVP, UAC, or
UVC, and X = invasive pressure
channel number 1 to 2.
Sensor
IP's not zeroed
LA/L lead off
LL/F lead off
RA/R lead off
RL/N lead off
Leads off
Move CO2 sensor to a cooler
location
Sensor overheating
2094480-001
Parameter
Temperature
ECG
ECG
ECG
ECG
Invasive pressures
SpO
2
Temperature
SpO
2
SpO
2
Invasive pressures
Invasive pressures
Invasive pressures
ECG
ECG
CO
2
CARESCAPE ONE
Alarm specifications
Alarm delay
7 s
7 s
10 s
<5 s
<5 s
<5 s
<5 s
<5 s
7 s
8 s
6 s
8 s
301 s (average) (300 to 301 s range)
LA/L lead off, LL/F lead off: 10 s
RA/R lead off: 11 s (average) (10 to 11
s range)
RL/N lead off: 16 s (average) (15 to 17
s range)
7 s
<5 s
271

Advertisement

Table of Contents
loading

Table of Contents