CARESCAPE ONE User Manual DRAFT 26 October 2017
SpO₂
Problem
Why is the parameter
window not displayed on
screen after connecting
the SpO
interface cable
2
and sensor?
152
Solution
It has been noted that letting the patient view the SpO
patient to assist in reducing motion artifact.
No SpO
data is displayed due to a hardware failure or an unrecognized or
2
defective sensor or cable.
Make sure the accessories are compatible.
●
Make sure the sensor is attached to the CARESCAPE SpO
●
CARESCAPE SpO
device is connected to the CARESCAPE ONE.
2
Disconnect and reconnect the sensor.
●
Disconnect and reconnect the cable.
●
Change the sensor.
●
Change the cable.
●
If the problem persists, contact qualified service personnel.
CARESCAPE ONE
waveform enables the
2
device and the
2
2094480-001