GE CARESCAPE ONE User Manual page 210

Hide thumbs Also See for CARESCAPE ONE:
Table of Contents

Advertisement

CARESCAPE ONE User Manual DRAFT 26 October 2017
Messages
Message
Artifact
Asystole
Bigeminy
Brady
HR LOW
HR(ECG) low
Check device
ECG malfunction
Couplet
Device overheating
Move ECG device to a
cooler location
ECG device failure:
Call service
ECG device
overheated. Shutting
down
210
Location
Possible causes
wavef.
Muscle artifact, high or low
frequency noise, or pacemaker
artifact.
al. area,
Physiological alarm.
param.,
wavef.
al. area,
Physiological alarm.
wavef.
al. area
Physiological alarm.
Measurement values are equal
to or outside the alarm limits.
wavef.
Hardware or software fault in
CARESCAPE ECG:
al. area
Cable fault
No updated data
Data sample drift
Configuration failure
al. area,
Physiological alarm.
wavef.
wavef.
CARESCAPE ECG surface
temperature is too high.
al. area
CARESCAPE ECG startup or
al. area
initialization failure.
al. area
CARESCAPE ONE turns off the
CARESCAPE ECG because its
surface temperature is too
high.
CARESCAPE ONE
Suggested actions
Check electrode contact.
Check lead placement.
Perform skin preparation.
Reposition/replace
electrodes.
Request the patient to remain
still.
Check if any electrical devices
are too close to the patient.
Check the patient status.
Check the patient status.
Check the patient status.
Adjust alarm limits if
necessary.
Reconnect the CARESCAPE
ECG.
Replace the CARESCAPE ECG.
If the problem persists,
contact qualified service
personnel.
Check the patient status.
Check the CARESCAPE ECG
placement and ensure
that it is not covered by
anything that might lead to
overheating.
Replace the CARESCAPE ECG.
Reconnect the CARESCAPE
ECG.
Replace the CARESCAPE ECG.
If the problem persists,
contact qualified service
personnel.
Check the CARESCAPE ECG
placement and ensure
that it is not covered by
anything that might lead to
overheating.
Replace the CARESCAPE ECG.
2094480-001