CARESCAPE ONE User Manual DRAFT 26 October 2017
Message
No battery backup
Patient admitted
Patient discharged
Powering down!
Replace battery
Service CS ONE
and specific error
indication
2094480-001
Location
Possible causes
al. area
There is no battery in the
CARESCAPE ONE.
al. area
ICU, NICU, and ED software
packages: The current patient
has just been admitted.
al. area
ICU, NICU, and ED software
packages: There is no admitted
patient.
al. area
The CARESCAPE ONE is being
used on battery power and there
is less than 1 minute of charge
left.
al. area
The battery is not working
properly.
al. area
Technical fault in the monitor.
CARESCAPE ONE
Messages
Suggested actions
Install a battery in the
CARESCAPE ONE battery
compartment.
No action required.
No action required.
Charge the battery by
connecting the CARESCAPE
ONE to a powered
CARESCAPE Dock F0.
Replace the battery.
Contact qualified service
personnel.