CARESCAPE ONE User Manual DRAFT 26 October 2017
Message
No battery backup
●
Patient admitted
●
Patient discharged
●
Powering down!
●
Replace battery
●
Service CS ONE
●
and specific error
indication
2094480-001
Location
Possible causes
al. area
There is no battery in the
●
CARESCAPE ONE.
al. area
ICU, NICU, and ED software
●
packages: The current patient
has just been admitted.
al. area
ICU, NICU, and ED software
●
packages: There is no admitted
patient.
al. area
The CARESCAPE ONE is being
●
used on battery power and there
is less than 1 minute of charge
left.
al. area
The battery is not working
●
properly.
al. area
Technical fault in the monitor.
●
CARESCAPE ONE
Messages
Suggested actions
Install a battery in the
●
CARESCAPE ONE battery
compartment.
No action required.
●
No action required.
●
Charge the battery by
●
connecting the CARESCAPE
ONE to a powered
CARESCAPE Dock F0.
Replace the battery.
●
Contact qualified service
●
personnel.
231