GE CARESCAPE ONE User Manual page 216

Hide thumbs Also See for CARESCAPE ONE:
Table of Contents

Advertisement

CARESCAPE ONE User Manual DRAFT 26 October 2017
Messages
Message
Incompatible probe
Incompatible SpO2
probe
Incompatible
adhesive SpO2 probe
Incompatible SpO2
device
Insufficient power for
SpO2 device
Low perfusion
Low signal
Low signal quality
Motion detected
Move SpO2 device to
a cooler location
No cable
No SpO2 cable
216
Location
Possible causes
param.
The sensor is not compatible or
it is not recognized.
al. area
Adhesive sensor is not
compatible or it is not
recognized.
al.area
CARESCAPE SpO
or initialization failure.
CARESCAPE SpO
hardware fault.
al. area
There is not enough power for
the CARESCAPE SpO
param.
Low perfusion at the
measurement point.
param.
The quality of the signal is
questionable.
param.
The quality of the signal is
questionable.
param.
Patient movement detected.
al. area
CARESCAPE SpO
temperature is too high.
param.
Interconnect cable is not
connected to the CARESCAPE
al. area
SpO
– Masimo.
2
CARESCAPE ONE
Suggested actions
Replace the sensor with a
compatible one. See the
supplemental information
provided.
If the problem persists,
contact qualified service
personnel.
device startup
Reconnect the CARESCAPE
2
SpO
2
device
Replace the CARESCAPE SpO
2
device.
If the problem persists,
contact qualified service
personnel.
Connect the acquisition
device.
platform to a host or use
2
mains power.
Charge the CARESCAPE ONE
battery.
Check the sensor and sensor
positioning.
Relocate the sensor to a
better measurement site, if
possible.
Make sure the patient is not
shivering.
Check the sensor placement
and the patient status.
Check the sensor and sensor
positioning.
Relocate the sensor to a
better measurement site, if
possible.
Make sure the patient is not
shivering.
Reposition sensor.
device surface
Check the CARESCAPE SpO
2
device placement and ensure
that it is not covered by
anything that might lead to
overheating.
Replace the CARESCAPE SpO
device.
Check the connection
between the interconnect
cable and CARESCAPE SpO
device.
device.
2
2
2
2
2094480-001