CARESCAPE ONE User Manual DRAFT 26 October 2017
Messages
Message
Incompatible probe
●
Incompatible SpO2
●
probe
Incompatible
●
adhesive SpO2 probe
Incompatible SpO2
●
device
Insufficient power for
●
SpO2 device
Low perfusion
●
Low signal
●
Low signal quality
●
Motion detected
●
Move SpO2 device to
●
a cooler location
No cable
●
No SpO2 cable
●
216
Location
Possible causes
param.
●
The sensor is not compatible or
it is not recognized.
al. area
●
Adhesive sensor is not
compatible or it is not
recognized.
al.area
CARESCAPE SpO
●
or initialization failure.
CARESCAPE SpO
hardware fault.
al. area
There is not enough power for
●
the CARESCAPE SpO
param.
●
Low perfusion at the
measurement point.
param.
●
The quality of the signal is
questionable.
param.
●
The quality of the signal is
questionable.
param.
●
Patient movement detected.
al. area
CARESCAPE SpO
●
temperature is too high.
param.
●
Interconnect cable is not
connected to the CARESCAPE
al. area
●
SpO
– Masimo.
2
CARESCAPE ONE
Suggested actions
Replace the sensor with a
●
compatible one. See the
supplemental information
provided.
If the problem persists,
●
contact qualified service
personnel.
device startup
Reconnect the CARESCAPE
●
2
SpO
2
device
Replace the CARESCAPE SpO
●
2
device.
If the problem persists,
●
contact qualified service
personnel.
Connect the acquisition
●
device.
platform to a host or use
2
mains power.
Charge the CARESCAPE ONE
●
battery.
Check the sensor and sensor
●
positioning.
Relocate the sensor to a
●
better measurement site, if
possible.
Make sure the patient is not
●
shivering.
Check the sensor placement
●
and the patient status.
Check the sensor and sensor
●
positioning.
Relocate the sensor to a
●
better measurement site, if
possible.
Make sure the patient is not
●
shivering.
Reposition sensor.
●
device surface
Check the CARESCAPE SpO
●
2
device placement and ensure
that it is not covered by
anything that might lead to
overheating.
Replace the CARESCAPE SpO
●
device.
Check the connection
●
between the interconnect
cable and CARESCAPE SpO
device.
device.
2
2
2
2
2094480-001