GE CARESCAPE ONE User Manual page 196

Hide thumbs Also See for CARESCAPE ONE:
Table of Contents

Advertisement

CARESCAPE ONE User Manual DRAFT 26 October 2017
CO₂
Problem
Why can we see dips in the capnogram during
expiration?
Why is the EtCO
value considerably lower than
2
the CO
partial pressure determined by blood gas
2
analysis?
The sample cell is inserted, but the pump does not
turn on and the Check Adapter message is displayed.
196
Solution
Dips in the capnogram can be due to patient
conditions. "Clefts" in the capnogram are seen
as action of muscle relaxant is impacted by
spontaneous ventilation.
The dips seen in the capnogram during expiration
may be related to the sidestream gas sampling, the
continuous gas flow to the Y-piece, and patient's
cardiac contractions, which cause intra-thoracic
pressure changes and therefore flow variations.
A pressure transducer in the LoFlo Sidestream
Module minimizes flow variations by measuring
and adjusting for changes in airway pressures.
EtCO
values are always lower than paCO
2
measured by blood gas analysis.
The typical clinical reasons are dead-space
ventilation, ventilation/perfusion mismatch, a drop
in cardiac output, alveolar shunts, and incomplete
emptying of the alveoli.
The EtCO
- paCO
2
a clinical condition such as an increase in alveolar
dead space.
Also check the integrity of the breathing
circuit; blood-gas analysis corrected to a lower
temperature in case of hypothermia.
Check that the sample cell is fully inserted.
The LoFlo Sidestream Module may need to warm
up to be detected.
Check that the LoFlo Sidestream Module is
connected to the CARESCAPE CO
Check the sample cell of the accessory for damage
or occlusion.
CARESCAPE ONE
values
2
gradient can be an indicator of
2
.
2
2094480-001

Advertisement

Table of Contents
loading

Table of Contents