CARESCAPE ONE User Manual DRAFT 26 October 2017
Messages
param. = parameter window
●
wavef. = waveform area
●
Message
Art 1 disconnect to
●
Art 2 disconnect
Disconnected
●
Art 1 sys high / Art 1
●
sys low to Art 2 sys
high / Art 2 sys low
Art 1 mean high / Art
●
1 mean low to Art
2 mean high / Art 2
mean low
Art 1 dia high / Art 1
●
dia low to Art 2 dia
high / Art 2 dia low
Artifact
●
Check device
●
Art X pressure
●
malfunction
CPP X pressure
●
malfunction
CVP X pressure
●
malfunction
Fem X pressure
●
malfunction
FemV X pressure
●
malfunction
ICP X pressure
●
malfunction
LAP X pressure
●
malfunction
PX pressure
●
malfunction
220
Location
Possible causes
al. area
No arterial invasive pressure is
●
detected.
param.
●
al. area
Measurement values are equal
●
to or outside the alarm limits.
param.
●
If Smart BP is enabled, this
is normal behavior when
zeroing, flushing, or sampling is
performed.
If Smart BP is not enabled, this
message indicates that the
measurement is disturbed by
artifact.
Systolic and diastolic numerics
are also invalidated if an artifact
event is detected while Smart BP
is enabled.
param.
●
The CARESCAPE Pressure has
malfunctioned.
al. area
●
CARESCAPE ONE
Suggested actions
Check the patient status.
●
Check connections.
●
If pressure drops because of
●
zeroing, perform the zeroing
process.
Check the patient status.
●
Adjust alarm limits if
●
necessary.
Check the patient status.
●
Check cable contact.
●
Minimize tubing length.
●
Rezero the transducer.
●
Check the patient status.
●
Check connections.
●
If the problem persists,
●
contact qualified service
personnel.
2094480-001