CARESCAPE ONE User Manual DRAFT 26 October 2017
Messages
param. = parameter window
wavef. = waveform area
Message
Art 1 disconnect to
Art 2 disconnect
Disconnected
Art 1 sys high / Art 1
sys low to Art 2 sys
high / Art 2 sys low
Art 1 mean high / Art
1 mean low to Art
2 mean high / Art 2
mean low
Art 1 dia high / Art 1
dia low to Art 2 dia
high / Art 2 dia low
Artifact
Check device
Art X pressure
malfunction
CPP X pressure
malfunction
CVP X pressure
malfunction
Fem X pressure
malfunction
FemV X pressure
malfunction
ICP X pressure
malfunction
LAP X pressure
malfunction
PX pressure
malfunction
220
Location
Possible causes
al. area
No arterial invasive pressure is
detected.
param.
al. area
Measurement values are equal
to or outside the alarm limits.
param.
If Smart BP is enabled, this
is normal behavior when
zeroing, flushing, or sampling is
performed.
If Smart BP is not enabled, this
message indicates that the
measurement is disturbed by
artifact.
Systolic and diastolic numerics
are also invalidated if an artifact
event is detected while Smart BP
is enabled.
param.
The CARESCAPE Pressure has
malfunctioned.
al. area
CARESCAPE ONE
Suggested actions
Check the patient status.
Check connections.
If pressure drops because of
zeroing, perform the zeroing
process.
Check the patient status.
Adjust alarm limits if
necessary.
Check the patient status.
Check cable contact.
Minimize tubing length.
Rezero the transducer.
Check the patient status.
Check connections.
If the problem persists,
contact qualified service
personnel.
2094480-001