CARESCAPE ONE User Manual DRAFT 26 October 2017
Messages
Message
CO2 measurement
removed
CO2 sensor removed
Sensor removed
CO2 zero required
Zero required
EtCO2 high
EtCO2 low
FiCO2 high
FiCO2 low
Identical CO2 device
Move CO2 device to a
cooler location
Device overheating
228
Location
Possible causes
al. area
The CARESCAPE CO
been disconnected from the
CARESCAPE ONE.
al. area
The LoFlo Sidestream Module
has been disconnected from the
param.
CARESCAPE CO
al. area
The CARESCAPE CO
zeroing.
param.
al. area
Measurement values are equal
to or outside the alarm limits.
al. area
Measurement values are equal
to or outside the alarm limits.
al. area
There is more than one
CARESCAPE CO
the system.
al. area
The CARESCAPE CO
temperature is too high.
param.
CARESCAPE ONE
Suggested actions
has
Connect the CARESCAPE CO
2
and LoFlo Sidestream Module
if you want to restart the
measurement.
Connect the LoFlo Sidestream
Module to the CARESCAPE
.
CO
2
2
measurement.
needs
Disconnect the patient and
2
perform zeroing from the
CO
2
room air.
If the problem persists,
replace the sampling line and
zero again.
Check the patient status.
Check the ventilator and
fresh gas flow settings.
Check arterial blood gas
values.
Adjust alarm limits if
necessary.
Check the patient status.
Check the ventilator or
absorber, breathing circuit,
and fresh gas flow settings.
Adjust alarm limits if
necessary.
Remove all but one
device in
CARESCAPE CO
2
surface
Check the CARESCAPE CO
2
placement and ensure
that it is not covered by
anything that might lead to
overheating.
Replace the CARESCAPE CO
Contact qualified service
personnel.
2
if you want to restart the
menu: Zeroing > Zero to
.
2
2
.
2
2094480-001