CARESCAPE ONE User Manual DRAFT 26 October 2017
AC mains to DC power supply for CARESCAPE Dock F0
The AC mains to DC power supply connects to the CARESCAPE Dock F0 power
receptacle and provides power to a docked CARESCAPE ONE. When connected to
the host monitor as an acquisition device, the host monitor provides power to the
CARESCAPE ONE, unless a 30 m host cable is used for the connection. In this case, the
host monitor does not provide power and the CARESCAPE ONE must be connected to
the AC mains to DC power supply for power.
Power supply mounting bracket
The power supply mounting bracket captures the power supply and allows the power
supply to be mounted to a roll stand.
Front and back views.
CARESCAPE Dock F0 side views
1
Docking interface connector to the CARESCAPE ONE.
2
ePort connector to the host monitor.