CARESCAPE ONE User Manual DRAFT 26 October 2017
Default settings
Setting
Description
Lead Analysis
Select the
lead(s) for
ECG analysis
when a 5-, 6-,
10-lead cable is
connected.
Show with HR
Select the
highest
deviation ST
value, PVC
value, a second
HR, or nothing
to display in the
ECG parameter
window.
ECG Grid
Display a grid
on waveforms.
ECG HR alarms default settings
Select Monitor Setup > Main Setup > Parameter Setup > Page 1 > ECG > HR Alarms
Arrhythmia > HR/PR to adjust these settings for the current patient when the HR Alarms
care unit setting is Multiple. The following table lists the factory default settings.
Setting
Description
HR alarm limits
Select the
high/low alarm
limits.
HR alarm on
Enable alarm
when limits are
violated.
352
ED
ICU
(3-lead cable connected: Single lead)
ED
ICU
150/50
(Neuro profile: 120/50;
Pediatric profile: 170/70;
Infant profile: 180/80)
(OR software package Perfusion profile: Disabled)
CARESCAPE ONE
NICU
OR
Multi lead
Off
Disabled
NICU
OR
200/90
160/40
(Infant
(Neuro
profile:
profile:
180/80)
120/50;
Adolescent
profile:
150/50;
Pediatric
profile:
170/70;
Infant
profile:
180/80)
Enabled
PACU
PACU
150/50
(Neuro
profile:
120/50;
Pediatric
profile:
170/70;
Infant
profile:
180/80)
2094480-001