Gas Specifications For Carescape Modules - GE auisys cs2 User's Reference Manual

Digital anesthesia carestation
Table of Contents

Advertisement

Aisys CS²

Gas specifications for CARESCAPE modules

Airway humidity
Sampling delay
Total system response time
Warm-up time
Respiration rate
Diversion flow
Airway pressure
CARESCAPE airway modules: Accuracy under different conditions
Ambient temperature: 10 to 40°C
Ambient pressure: 495 to 795 mmHg
Ambient Humidity: 10 to 98% RH, non-condensing
Automatic compensation for ambient pressure
Full module accuracy for Respiration rate of 4 to 70 breaths/min
During stable condition
CO2
+/- (0.2 vol% + 2% of reading)
O2
+/- (1 vol% + 2% of reading)
N2O
+/- (2 vol% + 2% of reading)
between 0 and 85 vol%
+/- (2 vol% + 8% of reading for
N2O between 85 and 100 vol%
Hal, Enf, Iso,
+/- (0.15 vol% + 5% of reading)
Sev, Des
11-20
.
CARESCAPE
0 to 100% condensing
3.0 seconds typical with a 3 m sampling line
Less than 3.8 seconds with a 3 m sampling line
1 minute for operation with CO2, O2, and N2O
5 minutes for operation of anesthetic agents
4 to 100 breaths/min
120 +/- 20 ml/min
-20 cmH2O to 100 cmH2O
.
During warm-up 1 to 10 minutes
(anesthetic agents 5-10 minutes)
+/- (0.4 vol% + 7% of reading)
+/- (3 vol% + 3% of reading)
+/- (3 vol% + 5% of reading)
+/- (0.3 vol% +10% of reading)
During warm-up 10-20 minutes
+/- (0.3 vol% + 4% of reading)
+/- (2 vol% + 2% of reading)
+/- (3% vol% + 3% of reading)
+/- (0.2 vol% +10% of reading
2067226-001

Advertisement

Table of Contents
loading

Table of Contents