CARESCAPE ONE User Manual DRAFT 26 October 2017
Setting
PR(Fem) Alarm On
UAC Sys High
Limit/Low Limit
UAC Sys Alarm On
UAC Mean High
Limit/Low Limit
UAC Mean Alarm On
UAC Dia High
Limit/Low Limit
UAC Dia Alarm On
PR(UAC) High
Limit/Low Limit
PR(UAC) Alarm On
Select Venous tab to adjust the invasive pressures venous alarm profile settings. The
following table lists the factory default settings.
2094480-001
ED
ICU
Enabled
(US only: Disabled)
90/40
(US only: NICU software package Enabled)
70/30
50/20
150/50
(Perfusion profile: Disabled)
CARESCAPE ONE
NICU
OR
Enabled
(Perfusion
profile:
Disabled)
90/40
(Cardiac,
Full-term, and
Infant profiles:
100/40)
(US only:
Infant profile:
100/50)
Disabled
60/30
(Cardiac,
Full-term, and
Infant profiles:
70/40)
(US only:
70/30; Infant
profile: 80/40)
Enabled
50/20
(Cardiac,
Full-term, and
Infant profiles:
60/20)
(US only:
Infant profile:
70/20)
Disabled
200/90
(Infant profile:
180/80)
Enabled
Default settings
PACU
Enabled
(US only:
Disabled)
90/40
70/30
50/20
150/50
335