Ecg Patient I/O Module Specifications; Sample Rate; Bandwidth; Isolation - GE Vivid i n Service Manual

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FL091021, R
IRECTION
5-8-2

ECG Patient I/O Module Specifications

5-8-2-1
Power Consumption
The ECG and Ext-ECG modules use 5 V (taken from the USB socket on the host).
Estimation: 80mA for ECG function and 50mA when working with EXT-ECG function.
5-8-2-2

Sample Rate

Sampling rate of the ECG signals is: 600 samples/second.
5-8-2-3

Bandwidth

The ECG Bandwidth is 0.05 to 150Hz.
5-8-3

Isolation

The ECG signals are isolated from the Vivid i n system to comply with IEC601-1 type CF device
classification. All leakage and auxiliary currents apply the values listed in
table IV in IEC601-1, section 19.3):
Table 5-12
Leakage Current
Earth leakage current
Enclosure leakage current
Patient leakage current
Patient auxiliary current
The ECG signals have double, reinforced isolation (B-a) from live parts. Hence, creepage distance and
air clearance in the module are 8 mm and 5 mm respectively, to withstand 4000 volts isolation between
applied part and grounded part.
The applied parts and enclosure have basic isolation (B-a).Therefore, creepage distance and air
clearance in the module are 4 mm and 2.5 mm respectively, to withstand 1500 volts isolation between
applied parts and enclosure.
5
EVISION
Leakage and Auxiliary Current Values
Normal Condition
0.5mA
0.1mA
0.01mA
0.01mA
Chapter 5 - Components and Function (Theory)
V
Table 5-12
Single Fault Condition
1mA
0.5mA
0.05mA
0.05mA
S
M
IVID I N
ERVICE
ANUAL
below (taken from
5-39

Advertisement

Table of Contents
loading

This manual is also suitable for:

H45561acH45551deFl000110Fl000090

Table of Contents