CARESCAPE ONE User Manual DRAFT 26 October 2017
Messages
Message
ICP 1 mean high / ICP
1 mean low to ICP
2 mean high / ICP 2
mean low
Identical pressure
devices
[Invasive pressure
channel label] X
sensor disconnected,
where [Invasive
pressure channel
label] = Art, CPP,
CVP, Fem, FemV, ICP,
LAP, P, PA, RAP, RVP,
UAC, or UVC, and X
= invasive pressure
channel number 1 to
2.
Sensor
[Invasive pressure
channel label] X
sensor failed, where
[Invasive pressure
channel label] = Art,
CPP, CVP, Fem, FemV,
ICP, LAP, P, PA, RAP,
RVP, UAC, or UVC, and
X = invasive pressure
channel number 1 to
2.
Sensor
PX not zeroed
where X = channel
number (1 or 2)
LAP 1 mean high / LAP
1 mean low to LAP
2 mean high / LAP 2
mean low
P1 standby to P2
standby
P1 mean high / P1
mean low to P2 mean
high / P2 mean low
222
Location
Possible causes
al. area
Measurement values are equal
to or outside the alarm limits.
al. area
There are two or more
CARESCAPE Pressures
connected.
al. area
The transducer detected a
disconnection.
param.
al. area
Faulty sensor or transducer.
param.
al. area
There is at least one invasive
pressure channel that has not
been zeroed.
al. area
Measurement values are equal
to or outside the alarm limits.
param.
The IP channel has been set to
standby.
al. area
Measurement values are equal
to or outside the alarm limits.
CARESCAPE ONE
Suggested actions
Check the patient status.
Adjust alarm limits if
necessary.
Remove all but one
CARESCAPE Pressure.
Check the patient status.
Check connections.
Acknowledge the alarm
if you are intentionally
disconnecting the invasive
pressure line.
Check cable and transducer
connections.
Replace the transducer.
Perform zeroing for all
channels.
Check the patient status.
Adjust alarm limits if
necessary.
Reactivate the channel by
selecting Activate P1 to
Activate P2.
Channels are reactivated if
pressures have remained
between 10 and 250 mmHg
for 10 seconds or longer.
Check the patient status.
Adjust alarm limits if
necessary.
2094480-001