CARESCAPE ONE User Manual DRAFT 26 October 2017
Alarm delay specifications for NIBP alarms
The following table lists the alarm delays for physiological alarms related to the non-invasive
blood pressure measurement. Any alarm delays that are less than 5 s are given as <5 s.
Alarm
Limit range
NIBP Sys high /
●
●
NIBP Sys low
NIBP Mean high /
●
●
NIBP Mean low
NIBP Dia high /
●
●
NIBP Dia low
Alarm delay specifications for invasive pressures alarms
The following table lists the alarm delays for physiological alarms related to the invasive
pressures measurement. Any alarm delays that are less than 5 s are given as <5 s.
NOTE
Alarm
Limit range
-25 to 320 mmHg
Art X sys high /
●
(-3.3 to 42.7 kPa)
Art X dia high /
Art X mean high
CVP X mean high
●
Fem X sys high /
●
Fem X dia high /
Fem X mean high
FemV X mean
●
high
ICP X mean high
●
LAP X mean high
●
P1 mean high to
●
P2 mean high
RAP X mean high
●
RVP X mean high
●
UAC X sys high /
●
UAC X dia high /
UAC X mean high
UVC X mean high
●
-25 to 320 mmHg
Art X sys low /
●
(-3.3 to 42.7 kPa)
Art X dia low /
Art X mean low
CVP X mean low
●
Fem X sys low /
●
Fem X dia low /
Fem X mean low
2094480-001
Limit increment
Adult: 25 to
Adult: 5 mmHg
260 mmHg
Child: 5 mmHg
Child: 25 to
Infant: 1 mmHg
190 mmHg
Infant: 15 to
140 mmHg
When an invasive pressure label is followed by an "X", the "X"
represents invasive pressure channel numbers 1 and 2.
Limit increment
1 mmHg (0.1 kPa)
1 mmHg (0.1 kPa)
CARESCAPE ONE
Alarm delay
<5 s
Alarm delay
Systolic: 13 s (average) (12 to 14 s range)
●
Diastolic: 13 s (average) (12 to 15 s range)
●
Mean: 14 s (average) (13 to 15 s range)
●
Systolic: 14 s (average) (14 to 16 s range)
●
Diastolic: 15 s (average) (14 to 15 s range)
●
Mean: 15 s (average) (14 to 15 s range)
●
Alarm specifications
267