CARESCAPE ONE User Manual DRAFT 26 October 2017
Alarm specifications
MESSAGE
NOTES
Unit setting HR
Brady
Alarms = Single;
Alarm priority =
Escalating; HR
<0.75 x low
limit; Primary
HR source AUTO,
IntelliRate, or ECG.
Brady
Unit setting HR
Alarms = Single;
non-ECG source
and Primary HR
source AUTO,
IntelliRate.
Brady
Unit setting HR
Alarms = Single;
Alarm priority =
Escalating;
non-ECG source
and Primary HR
source AUTO,
IntelliRate.
Brady
Unit setting HR
Alarms = Single;
Alarm priority =
Escalating; HR
<0.75 x low limit;
non-ECG source
and Primary HR
source AUTO,
IntelliRate.
Check device
Couplet
Device
overheating
ECG device
failure: Call
service
ECG device
overheated.
Shutting down
ECG malfunction
274
LOCATION
PRIORITIES AND ESCALATION TIMES
info
alarm
alarm
alarm
alarm
0 s
wavef.
alarm,
0 s , according to priority setting: high, medium, low,
wavef.
off, info
0 s
wavef.
alarm
alarm
alarm
CARESCAPE ONE
low
medium
High, medium, or low according to
the priority setting. The alarm delay is
equivalent to the set PR alarm delay,
which is configurable from 0 to 20 s in
increments of 5 s.
Accord-
ing to the
set PR
alarm de-
lay, which
is config-
urable from
0 to 20 s in
increments
of 5 s.
0 s
0 s
0 s
high
0 s
Alarm delay
+ 64 s
Accord-
ing to the
set PR
alarm de-
lay, which
is config-
urable from
0 to 20 s in
increments
of 5 s.
2094480-001