CARESCAPE ONE User Manual DRAFT 26 October 2017
Alarm specifications
Alarm
Move SpO2 device to a cooler
location
NIBP cuff loose
NIBP cuff occlusion
Cuff occlusion
NIBP auto stopped
NIBP STAT stopped
NIBP over range
NIBP under range
Noise
Noisy ECG
Pressure device failure: Call
service
Pressure device initializing
Pressure device overheated.
Shutting it down
Pressure measurement
removed
Resp (Imped.) measurement
paused
Measurement paused
Sample line disconnected
Line disconnected
SpO2 device failure: Call
service
SpO2 device overheated
Shutting down
SpO2 faulty probe
Faulty adhesive SpO2 probe
SpO2 measurement removed
Temp device failure: Call
service
Temp device overheated.
Shutting down
Temp measurement removed
272
Parameter
SpO
2
NIBP
NIBP
NIBP
NIBP
ECG
Invasive pressures
Invasive pressures
Invasive pressures
Invasive pressures
Impedance respiration
CO
2
SpO
2
SpO
2
SpO
2
SpO
2
Temperature
Temperature
Temperature
CARESCAPE ONE
Alarm delay
6 s
22 s (average) (22 s range)
<5 s
<5 s
<5 s
27 s (average) (25 to 27 s range)
7 s
6 s
10 s
<5 s
50 s (average) (49 to 51 s range) for all
software packages except NICU.
NICU software package: 18 s (average)
(17 to 20 s range)
42 s (average) (41 to 42 s range)
6 s
11 s (average) (9 to 12 s range)
7 s
<5 s
8 s
8 s
<5 s
2094480-001