CARESCAPE ONE User Manual DRAFT 26 October 2017
Default settings
Setting
CVP Mean High
Limit/Low Limit
CVP Mean Alarm On
UVC Mean High
Limit/Low Limit
UVC Mean Alarm On
FemV Mean High
Limit/Low Limit
FemV Mean Alarm On
Select PA, ICP & P1-P2 tab to adjust the invasive pressures alarm profile settings. The
following table lists the factory default settings.
Setting
PA Sys High Limit/Low
Limit
PA Sys Alarm On
PA Mean High
Limit/Low Limit
PA Mean Alarm On
PA Dia High Limit/Low
Limit
PA Dia Alarm On
Neuro profiles:
ICP Mean High
Limit/Low Limit
ICP Mean Alarm On
CPP High Limit/Low
Limit
CPP Alarm On
336
ED
ICU
15/0
(Infant profile: 10/0)
Enabled
Enabled
(Trauma and
Neuro profiles:
Disabled)
15/0
(Infant profiles: 10/0)
ED
ICU
Enabled
(Trauma and
Disabled)
15/0
(Infant profile: 9/0)
100/60
(Infant profile: 90/50)
CARESCAPE ONE
NICU
OR
5/0
(Infant profile: 10/0; Perfusion
Enabled
15/0
Disabled
5/0
(Infant profiles: 10/0)
Enabled
NICU
OR
40/10
Disabled
(US only: Enabled)
30/5
Enabled
20/5
Enabled
9/0
(Infant profile: 9/0)
Enabled
100/40
(Infant profile: 90/50)
Enabled
PACU
15/0
profile: 15/-5)
15/0
PACU
15/0
100/60
2094480-001