CARESCAPE ONE User Manual DRAFT 26 October 2017
Message
ECG measurements
removed
Frequent PVCs
Frequent SVCs
Tachy
HR high
HR(ECG) high
Identical ECG device
Irregular
Lead off
LA/L lead off
LL/F lead off
RA/R lead off
RL/N lead off
V2/C2 lead off
V3/C3 lead off
V4/C4 lead off
V5/C5 lead off
V6/C6 lead off
V/C lead off
Va/Ca lead off
Vb/Cb lead off
Lead changed
Leads off
Learning
Missing Beat
2094480-001
Location
Possible causes
al. area
CARESCAPE ECG has been
removed.
al. area
Physiological alarm.
al. area,
Physiological alarm.
wavef.
al. area
Measurement values are equal
to or outside the alarm limits.
Physiological alarm.
al. area
There are two CARESCAPE ECGs.
al. area,
Physiological alarm.
wavef.
al. area,
One or more leadwires or
wavef.
electrodes disconnected. Other
ECG leads are available for
arrhythmia detection.
wavef.
The monitor automatically
switches the ECG1 waveform
selection to a measurable ECG
Lead (I, II, III, aVR, aVL, aVF,
V1, V2, V3, V4, V5 or V6) if the
current ECG1 waveform is not
measurable. This message also
appears when the user changes
the ECG1 lead selection.
al. area,
One or more of the connected
wavef.
leadwires is disconnected and
arrhythmia detection is not
possible.
wavef.
ECG algorithm is in learning
phase, message shown e.g.
when ECG measurement is
started.
al. area,
Physiological alarm.
wavef.
CARESCAPE ONE
Messages
Suggested actions
Check all connections and
reconnect as required.
Check the patient status.
Check the patient status.
Check the patient status.
Adjust alarm limits if
necessary.
Remove one CARESCAPE ECG.
Check the patient status.
Check the leadwire and
electrode connections.
Note that the ECG waveform
will be from the lead selected
for the ECG1 lead. Check the
lead.
Check the connections.
No action required.
Check the patient status.
211