CARESCAPE ONE User Manual DRAFT 26 October 2017
Message
ECG measurements
●
removed
Frequent PVCs
●
Frequent SVCs
●
Tachy
●
HR high
●
HR(ECG) high
●
Identical ECG device
●
Irregular
●
Lead off
●
LA/L lead off
●
LL/F lead off
●
RA/R lead off
●
RL/N lead off
●
V2/C2 lead off
●
V3/C3 lead off
●
V4/C4 lead off
●
V5/C5 lead off
●
V6/C6 lead off
●
V/C lead off
●
Va/Ca lead off
●
Vb/Cb lead off
●
Lead changed
●
Leads off
●
Learning
●
Missing Beat
●
2094480-001
Location
Possible causes
al. area
CARESCAPE ECG has been
●
removed.
al. area
Physiological alarm.
●
al. area,
Physiological alarm.
●
wavef.
al. area
Measurement values are equal
●
to or outside the alarm limits.
Physiological alarm.
al. area
There are two CARESCAPE ECGs.
●
al. area,
Physiological alarm.
●
wavef.
al. area,
One or more leadwires or
●
wavef.
electrodes disconnected. Other
ECG leads are available for
arrhythmia detection.
wavef.
The monitor automatically
●
switches the ECG1 waveform
selection to a measurable ECG
Lead (I, II, III, aVR, aVL, aVF,
V1, V2, V3, V4, V5 or V6) if the
current ECG1 waveform is not
measurable. This message also
appears when the user changes
the ECG1 lead selection.
al. area,
One or more of the connected
●
wavef.
leadwires is disconnected and
arrhythmia detection is not
possible.
wavef.
ECG algorithm is in learning
●
phase, message shown e.g.
when ECG measurement is
started.
al. area,
Physiological alarm.
●
wavef.
CARESCAPE ONE
Messages
Suggested actions
Check all connections and
●
reconnect as required.
Check the patient status.
●
Check the patient status.
●
Check the patient status.
●
Adjust alarm limits if
●
necessary.
Remove one CARESCAPE ECG.
●
Check the patient status.
●
Check the leadwire and
●
electrode connections.
Note that the ECG waveform
●
will be from the lead selected
for the ECG1 lead. Check the
lead.
Check the connections.
●
No action required.
●
Check the patient status.
●
211