CARESCAPE ONE User Manual DRAFT 26 October 2017
Message
Check CO2 adapter
●
Check Adapter
●
Check sample line
●
CO2 device failure:
●
Call service
CO2 device initializing
●
CO2 device
●
malfunction
Check device
●
CO2 device
●
overheated Shutting
down
2094480-001
Location
Possible causes
al. area
There is an obstruction in
●
the LoFlo receptacle sample
param.
●
cell, or the LoFlo receptacle
sample cell is not within the
optimal operating range for CO
measurement.
al. area
Sample flow rate is lower or
●
higher than the nominal flow
param.
●
due to an obstruction in the
sample line or exhaust port, or a
cut or split in the sample line.
The filter in the sample line is
completely saturated.
If the situation persists for more
than 2 minutes, the pump will
turn off.
CARESCAPE CO
al. area
●
initialization failure.
al. area
The CARESCAPE CO
●
software download mode.
CARESCAPE CO
al. area
●
runtime failure.
param.
●
al. area
The CARESCAPE CO
●
temperature is too high and
the CARESCAPE ONE turns the
measurement off.
CARESCAPE ONE
Suggested actions
Ensure that the sampling line
●
connector is correctly and
fully inserted.
Check that the sample cell
●
2
windows and receptacle
windows do not have any
foreign material on them.
Perform zeroing procedure to
●
optimize the operation.
If the problem persists,
●
replace the adapter.
Check the sample line and
●
exhaust port for obstructions.
Check the sample line for
●
cuts or splits and replace if
necessary.
Remove the sample line
●
and dispose of it following
hospital protocol.
If the pump has shut off,
●
disconnect then reconnect
the accessory to restart the
pump
startup or
Reconnect the CARESCAPE
●
2
CO
2
Replace the CARESCAPE CO
●
Contact qualified service
●
personnel.
is in the
Wait and see if the message
●
2
disappears.
Replace the CARESCAPE CO
●
Contact qualified service
●
personnel.
startup or
Reconnect the CARESCAPE
●
2
CO
2
Replace the CARESCAPE CO
●
Contact qualified service
●
personnel.
surface
Check the CARESCAPE CO
●
2
placement and ensure
that it is not covered by
anything that might lead to
overheating.
Replace the CARESCAPE CO
●
Contact qualified service
●
personnel.
Messages
.
.
2
.
2
.
.
2
2
.
2
227