CARESCAPE ONE User Manual DRAFT 26 October 2017
Message
Check CO2 adapter
Check Adapter
Check sample line
CO2 device failure:
Call service
CO2 device initializing
CO2 device
malfunction
Check device
CO2 device
overheated Shutting
down
2094480-001
Location
Possible causes
al. area
There is an obstruction in
the LoFlo receptacle sample
param.
cell, or the LoFlo receptacle
sample cell is not within the
optimal operating range for CO
measurement.
al. area
Sample flow rate is lower or
higher than the nominal flow
param.
due to an obstruction in the
sample line or exhaust port, or a
cut or split in the sample line.
The filter in the sample line is
completely saturated.
If the situation persists for more
than 2 minutes, the pump will
turn off.
CARESCAPE CO
al. area
initialization failure.
al. area
The CARESCAPE CO
software download mode.
CARESCAPE CO
al. area
runtime failure.
param.
al. area
The CARESCAPE CO
temperature is too high and
the CARESCAPE ONE turns the
measurement off.
CARESCAPE ONE
Suggested actions
Ensure that the sampling line
connector is correctly and
fully inserted.
Check that the sample cell
2
windows and receptacle
windows do not have any
foreign material on them.
Perform zeroing procedure to
optimize the operation.
If the problem persists,
replace the adapter.
Check the sample line and
exhaust port for obstructions.
Check the sample line for
cuts or splits and replace if
necessary.
Remove the sample line
and dispose of it following
hospital protocol.
If the pump has shut off,
disconnect then reconnect
the accessory to restart the
pump
startup or
Reconnect the CARESCAPE
2
CO
2
Replace the CARESCAPE CO
Contact qualified service
personnel.
is in the
Wait and see if the message
2
disappears.
Replace the CARESCAPE CO
Contact qualified service
personnel.
startup or
Reconnect the CARESCAPE
2
CO
2
Replace the CARESCAPE CO
Contact qualified service
personnel.
surface
Check the CARESCAPE CO
2
placement and ensure
that it is not covered by
anything that might lead to
overheating.
Replace the CARESCAPE CO
Contact qualified service
personnel.
Messages
.
.
2
.
2
.
.
2
2
.
2
227