CARESCAPE ONE User Manual DRAFT 26 October 2017
Alarm specifications
Alarm
Check device
Art X pressure malfunction
CPP X pressure malfunction
CVP X pressure malfunction
Fem X pressure malfunction
FemV X pressure malfunction
ICP X pressure malfunction
LAP X pressure malfunction
PAX pressure malfunction
RAP X pressure malfunction
RVP X pressure malfunction
UAC X pressure malfunction
UVC X pressure malfunction
where X = channel number (1 or
2)
Check device
SpO2 malfunction
Check device
Temperature malfunction
Check NIBP
Check sample line
Check sensor
Service CO2 sensor
Check SpO2 probe
Check probe
CO2 device failure: Call service
CO2 device overheated
Shutting down
CO2 sensor removed
Sensor removed
CO2 zero required
Zero required
Device overheating
Move CO2 device to a cooler
location
Device overheating
Move ECG device to a cooler
location
Device overheating
Move pressure device to a
cooler location
270
Parameter
Invasive pressures
SpO
2
Temperature
NIBP
CO
2
CO
2
SpO
2
CO
2
CO
2
CO
2
CO
2
CO
2
ECG
Invasive pressures
CARESCAPE ONE
Alarm delay
<5 s
<5 s
8 s
<5 s
40 s (average) (39 to 40 s range)
<5 s
<5 s
7 s
10 s
<5 s
42 s (average) (41 to 42 s range)
6 s
6 s
7 s
2094480-001